Fuel Cell Analyzer As-510-4 |
|
Capable of evaluatiing fuel cells fitted with many types of resference electrodes.
Supports detailed evaluation and analysis of anode and cathode characteristics. |
|
|
 |
Features |
 |
|
|
 |
Equipped with input terminals for three-terminal voltage measurement (+, -, and REF), the As-510-4 can simultaneously measure the voltage between terminals and isolated resistance (IR). It incorporates new circuitry to provide support for a wide variety of reference electrode types.
|
 |
NFs exclusive step method supports IR measurement with
high reproducibility, while also supporting the current inter-
ruption method.
|
 |
When combined with the optional three-terminal impedance measuring unit, the As-510-4 can measure impedance between terminals and supports continuous measurements without having to change connections.
|
 |
Three ranges (10A, 50A, and 100A), and capacity of up to
20V or 100W.
|
 |
More precise current and voltage settings and more accurate
measurement provide improved reproducibility.
|
 |
Compatible with constant-current/constant-voltage operation,
and supports Open Circuit Voltage (OCV) measurement.
|
 |
Equipped with a GPIB interface.
|
 |
Main Specifications |
 |
|
|
Voltage: 0V to 20V
Power: 100W
Current: 10A (low), 50A (mid), and 100A (high)
Current setting accuracy: ±0.05% of full scale max.
(constant-current operation in all ranges)
Current measurement accuracy: ±0.05% of full scale max. (auto range)
Voltage setting accuracy: ±0.5% of set value + ±0.2% of full scale max.
Voltage measurement accuracy: ±0.1% of full scale max.
IR measurement range: 0.0000m&Mac189; to 999.99m&Mac189; (five-digit display)
IR measurement timing: Variable, between 10µs and 1000µs after interruption
Impedance frequency measurement range: 0.1mHz to 50kHz (full range) |
|