冷阴极放电管用特性测试装置
冷阴极放电管用特性测试装置
As-114B / As-128 / As-130

System Structure

Characteristics tester for cold cathode fluorescent lamps (Main unit) ..... 1 unit
Test bench for cold cathode fluorescent lamps (CCFL test bench) .......... 1 unit
Accessory kit (CCFL clip terminals, connection cable (3m) between main unit and CCFL test bench, test bench high voltage cable, etc.)

Main Unit

Test Bench

Power Supply Section

Output Frequency
25kHz to 100kHz, Accuracy : ±1% of set value
0.1kHz step variable, sine wave
External Signal Input*
25kHz to 100kHz, sine wave
Output Voltage As-114 B : 3kVrms(high on one end), 3kVrms(high on both ends)
As-128 : 4kVrms(high on one end), 4kVrms(high on both ends)
As-130 : 4kVrms(high on one end), 6kVrms(high on both ends)
Output Current 20mArms
Output Voltage
Distortion Factor
Within 5% (25kHz to 100kHz)
CV Mode Constant voltage mode that controls the output voltage Vo at a constant value.
CC Mode
Constant current mode that controls the lamp current I at a constant value.
Vds Mode The mode that measures the lighting start voltage and discharge time log.

Voltage Measurement Section (Vo, V)

AC Monitor
Output
Gain 1/1000
Accuracy ±3% of the full scale*1
Indicator Display Range
As-114B : 0 to 3000Vrms, 0 to 9999Vp-p
As-128: 0 to 4000Vrms, 0 to 9999Vp-p
As-130: 0 to 6000Vrms, 0 to 9999Vp-p
Accuracy ±3% of the full scale

Current Measurement Section (I)

AC Monitor
Output
Gain 1Vrms/10mArms
Accuracy ±3% of the full scale
Indicator Display Range
0 to 20.00mArms
Accuracy ±3% of the 20mArms

CCFL Test Bench

Acceptable Lamp*2
Dimension
As-114B : 50mm to 1200mm
As-128: 50mm to 1500mm
As-130: 50mm to 1500mm
External Inverter
Connection
Voltage and current can be measured by using an
external inverter.

General Items

Power Supply
AC 100V(AC110V, 200V or 220V is available) ±10%,
48Hz to 62Hz
Power Consumption As-114B : 600VA or lower, As-128 : 800VA or lower,
As-130: 1000VA or lower
Operating
Temperature/
Humidity
Main Unit
0 to +40ºC, 10 to 85% RH
CCFL Test
Bench
-25ºC*3 to +105ºC, 0 to 95% RH
Maximum wet bulb temperature: 38ºC
GPIB Interface
Reads Vo, VR, and I values,
Switches among CV, CC, and Vds modes,
Switches V rms/p-p, turns on/off high voltage output
Shock Guard Function
The high voltage output is cut off when the CCFL test
bench cover is opened.
Withstand Voltage Between power input pairs and others: AC 1500V,
1 minute
Insulating Resistance Between power input pairs and others: 50Mω or
higher at DC 500V
*1 : The full scale value : 3Vrms with As-114B, 4Vrms with As-128, 6Vrms with As-130.
The output voltage is calibrated with our High voltage divider.
*2 : CCFL test bench can be manufactured according to customer's request of CCFL's length and shape.
*3 : We can also make a test bench that functions at -30ºC.